To set up a Beckman DXH520 a Static IP address and port number would need to be assigned for the instrument.

  • On the instrument settings find settings then click LIS setup, it will appear as bellow
  • This is where the IP address and the port number goes and 
  • Protocol should be TCP/IP



On preLink


  • Go to Administration then Instrument management > Add new record 
  •  A window like the one below will appear 


  • The Analyzer IP address and port number should be the same as the one on the instrument. 
  • Click update then restart the Instrument Control App for the changes to take effect.